Appendix C: Data Port Protocol
The analog function can be accessed from the main menu by
pressing SETUP, NEXT, NEXT, ANALOG
. Selecting “0 VOLT”
or “1 VOLT” causes that voltage to appear at pins 6, 3, or 10
.
Selecting “STEP” causes the voltage to increase from 0 to 1 volt
at 1/10th second increments, with each step lasting at least 1
second.
Calibration of the analog output and the attached device can be
performed by qualified service personnel using the procedure
described in the NPB-295 service manual.
81