Start-Up and Use
Ensure that the supply voltage selector switch on the rear panel is
set to the proper voltage.
Power-On Self-Test (POST)
WARNING: Ensure that the speaker is clear of any
obstruction. Failure to do so could result in an inaudible
alarm tone.
1. Plug an appropriate Nellcor sensor firmly into the MC-10 or
SCP-10 and lower the MC-10 or SCP-10 sensor lock over
the sensor connector until it clicks into place. Insert the MC-
10 or SCP-10 into the NPB-295 sensor port. Apply the
sensor to the patient as described in the sensor directions for
use.
2. Turn on the NPB-295 by pressing the Power On/Off Button.
The monitor automatically starts a power-on self-test
(POST), which tests its circuitry.
3. During the POST, the entire display lights and then the
Nellcor logo with model number and software version are
displayed for approximately 3 seconds. All indicator lights
illuminate briefly.
Caution: If any indicator or display element does not light,
do not use the monitor. Instead, contact qualified service
personnel, your local Mallinckrodt representative, or the
Mallinckrodt’s Technical Services Department.
4. If the NPB-295 detects an internal problem during POST, an
error code or error message may be displayed and a low
priority alarm will sound. Depending on the reason for the
failure, the screen may be blank or the low priority alarm
may not sound. Refer to the Troubleshooting section for a
list of correctable error messages.
5. Upon successful completion of the POST, the NPB-295
sounds a 1-second tone indicating that the monitor has
passed the test.
WARNING: If you do not hear the POST pass tone, do not
use the monitor.
24