Sensors and Accessories
PERFORMANCE CONSIDERATIONS
WARNING: Pulse oximetry readings and pulse signal can be
affected by certain ambient environmental conditions, sensor
application errors, and certain patient conditions.
Inaccurate measurements can be caused by:
incorrect application of the sensor
placement of the sensor on an extremity with a blood
pressure cuff, arterial catheter, or intravascular line
• ambient light
excessive patient movement
• venous pulsations
intravascular dyes, such as indocyanine green or methylene
blue
• defibrillation
Other physiological conditions or medical procedures that may
interfere with the monitor’s measurements include significant
levels of dysfunctional hemoglobin, low perfusion, and dark
pigment.
Loss-of-pulse signal can occur for the following reasons:
the sensor is too tight
a blood pressure cuff is inflated on the same extremity as the
one with the sensor attached
there is arterial occlusion proximal to the sensor
Select an appropriate sensor, apply it as directed, and observe all
warnings and cautions presented in the directions for use
accompanying the sensor. Clean and remove any substances such
as nail polish from the application site. Periodically check to
ensure that the sensor remains properly positioned on the patient.
19