Start-Up and Use
ADJUSTABLE SETTINGS
The following adjustments can be made using the Adjust
Up/Down and Alarm Silence Buttons.
Pulse beep volume
• Alarm volume
Alarm silence duration
Disabling audible alarms
Pulse Beep Volume
To adjust the pulse beep volume during normal monitoring, press
and hold the Adjust Up or Adjust Down Button to change the
setting. Pressing and holding the Adjust Down Button will cause
the volume to decrease until it is no longer heard.
Alarm Volume
To view the current volume of the audible alarm, press and hold
the Alarm Silence Button for more than 3 seconds. The current
volume level is indicated in the Pulse Rate Display as a value
from 1 (lowest) to 10 (highest). A tone at the displayed level
sounds.
To adjust the volume, press and hold the Alarm Silence Button
for more than 3 seconds, then press the Adjust Up or Adjust
Down Button to change the setting. The volume cannot be set to
zero.
Alarm Silence Duration
Alarms can be silenced for a preset period called the audible
alarm silence duration. To view the current setting, press and
hold the Alarm Silence Button for less than 3 seconds. To adjust
the setting, press and hold the Alarm Silence Button (for less than
3 seconds) and use the Adjust Up or Adjust Down Buttons to
increase or decrease the value. Possible values are 30, 60, 90, or
120 seconds, or OFF. (The OFF selection is discussed later in
this section.)
The audible alarm silence duration begins when the Alarm
Silence Button is pressed (for less than 3 seconds) and released.