Mallinckrodt NPB-190 Manuel de l'opérateur Page 57

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 87
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 56
Troubleshooting and Maintenance
DISALLOWED ON BATTERY - An attempt to print or
download data port information while operating on battery power
has been made. Connect to AC power and retry.
DISALLOWED ON LOW BATTERY - An attempt to turn on
the backlight has been made while in a low battery condition. If
the backlight is turned off during a low battery condition, it
cannot be turned back on.
READING TRENDS - The monitor is gathering trend
information for display.
INVALID SILENCE DURATION - An attempt has been made to
set the alarm silence duration power-on default to “OFF”. The
power-on default cannot be set to “OFF”.
INVALID SPO2 LIMIT - An attempt has been made to set either
the upper or lower alarm limit power-on default below 80. The
power-on default cannot be set below 80.
Suggested Corrective Actions
If you experience a problem while using the NPB-295 and are
unable to correct it, contact qualified service personnel or your
local Mallinckrodt representative. The NPB-295 service manual,
which is for use by qualified service personnel, provides
additional troubleshooting information.
Following is a list of possible errors and suggestions for
correcting them.
1. There is no response to the Power On/Off Button.
If operating on AC power, ensure that the supply voltage
selector switch is set to the proper voltage.
If operating on AC power, the fuse may be blown.
Notify service personnel to check and, if necessary,
replace the fuse.
If operating on battery power, the battery may be
missing or discharged. If the battery is discharged,
notify service personnel to charge or replace the battery.
51
Vue de la page 56
1 2 ... 52 53 54 55 56 57 58 59 60 61 62 ... 86 87

Commentaires sur ces manuels

Pas de commentaire